UT-PTP7 Rev. 03 Page 12 of 14 07/01/2021
c. Corrosion signals with substantial change in depth likely have responses that move
across a significant amount of the time base.
d. Very unlikely to have amplitude larger than the back-wall. If wall loss is very deep
(close to scanning surface) and has a large plateau (flat area) then it is possible to have
a larger amplitude than a back-wall response, but this is very rare.
e. Corrosion is very unlikely to create multiple, or repeat, signals. If wall loss is very deep
(close to scanning surface) and has a large plateau (flat area) then it is possible, but
this is also very rare. Samples with erosion that have a smooth area that is parallel to
the scanning surface are an exception to this guidance. These types of areas can
generate multiple reflection signals like a lamination. However, areas of erosion will
generate changes in signal behavior when moving the transducer on and off of the
eroded area in various scanning directions unlike laminations which will generate the
same signal behavior in every direction as the transducer is moved on and off of the
area containing the lamination.
f. Shallow internal pitting can be seen in small changes to the back-wall signal. Monitor
drops in signal amplitude in conjunction with slight movement to the left on the time
base line.
g. Shallow pitting and corrosion will generate signals close to the full sample back-wall
thickness at undamaged areas.
h. Isolated areas of corrosion with steep sides to the wall loss (i.e. large single pits) can
generate A-scan signals similar to large laminations where the back-wall signal
disappears completely, and a signal appears with shorter sound path. In a situation
where this occurs, care must be taken to verify if there is any movement on the time
base to the left (shorter sound path) along with reduced amplitude of the back-wall
signal and/or a slight movement to the left (shorter sound path) along with increasing
amplitude of the mid-wall signal. If either of these conditions occur the reflector is
likely an isolated pit with significant wall loss rather than a mid-wall lamination.
3. Coated parts can cause additional difficulties in signal evaluation:
a. Coating can cause an increase in signal ring-down, additional cycles in the sine wave,
which makes all signals seem broader with an increased number of signal cycles seen
on the A-scan. This makes all signals, including clean back-wall signals and
laminations, to appear broader. Coating on samples that creates additional signal
ring-down can also cause laminations to not generate multiple signals.
b. Extra care must be taken on coating parts when evaluating signals from laminations
to verify they do not move back and forth on the time base.